More and more hospitals are considering adding PET scanners to
their nuclear medicine imaging arsenal. And a variety of sites are taking a long, hard
look at the prospect of PET-CT. As with any capital purchase, there are a number of
variables to consider. How much will the equipment cost? Does the facility have the
patient volume necessary to justify the purchase of a PET scanner? Does the hospital have
the necessary in-house expertise to staff the new equipment? If not, is it possible to
train or attract outsiders to the site? And how does an on-site cyclotron fit into the PET
equation? A final variable that should be inherent in each phase of the decision-making
process is radiation safety. How does the facility intend to insure the safety of PET
technologists and patients?
Standard nuclear medicine radiation safety practices are a good place to start, but
they are not the final answer for developing PET safety protocols. Kelly Classic, medical
health physicist and administrator for radiation safety at Mayo Clinic (Rochester, Minn.),
says, Initially, vendors claimed PET safety and shielding requirements were not any
different from standard nuclear medicine safety and shielding requirements. And it seemed
to make sense at first. But the reality of PET scanning is that it is in a class by
itself as far as radiation safety and shielding are concerned. Christina Truelock,
assistant chief technologist of nuclear medicine at Thomas Jefferson University Hospital
(Philadelphia), confirms, A PET scanner does require safety equipment and protocols
above and beyond a what is used in a routine nuclear medicine department. A PET scanner
needs heavier shielding, and technologists need to adapt their handling and radiation
safety techniques.
Still, the basic principles of PET safety and shielding do have their roots in good
old-fashioned nuclear medicine safety and shielding. Scott Holbrook, chair of the PET
Education Task Force for the technologist section of the Society of Nuclear Medicine
(Reston, Vir.), explains, Good radiation safety practices apply to PET and nuclear
medicine in the same way. The same principles of physics apply. However, there are
few key distinctions between PET and nuclear medicine safety protocols; these distinctions
represent a primary concern for facilities jumping into PET.
The quick tour of nuke med safety protocols
Standard nuclear medicine radiation safety practices boil down to three
considerations-time, distance and shielding. Technologists need to limit their time of
exposure to radiation sources and pay attention to their distance from sources and
hospitals need to install and maintain proper shielding. Direct handling of
radiopharmaceuticals is completed quickly, and techs are trained not to sit a patient or
spend an inordinate amount of time with a patient. Regardless of training, techs are
caregivers and Holbrook admits, In standard nuclear medicine, it is not uncommon for
techs to sit or stand the next to the patient for reassurance after injecting the
radioisotopes is generally accepted. The rationale is simple; it is somewhat
acceptable for techs to be close to the patients because they are giving psychological
care as well as completing the exam and most of the doses from patients for general
diagnostic are very, very low.
Whether or not techs opt to hold patients hands, the principles of time, distance
and shielding do keep most nuclear medicine technologists safe from overexposure. Truelock
notes, In general nuclear medicine, the exposure rate is minimal. Technologists
generally dont get overexposed.
The National Council on Radiation Protection and Measurements (Bethesda, Md.) estimates
that the average nuclear medicine technologists whole body exposure is about 400
mrem annually. Mayo Clinic separately monitors its PET and nuclear medicine technologists
and reports that PET techs receive about three times the radiation dosage of nuclear
medicine technologists. That figure may be even higher for inexperienced PET
technologists. Holbrook says, Techs will experience increased radiation exposure as
they first move into PET, but a few months after beginning to work in PET it comes down to
a more acceptable level. The key point for any facility considering moving into PET
is to start with a rigorous radiation safety protocol.
PET safety protocols
How can a facility insure the safety of its staff as it implements a PET scanner? The same
general principles of time, distance and shielding that apply in nuclear medicine are used
but in a much more stringent manner. The difference is a matter of simple physics. The
gamma emitters used for diagnostic nuclear medicine are actually fairly low-energy
135 to 150-keV, which means there isnt much exposure from the patient. PET, however,
is a whole different ball of wax.
Mary Anne Dell, vice president of manufacturing and health physicist for Capintec, Inc.
(Pittsburgh, Pa.), explains, All PET isotopes, by definition, are positron emitters
and have an energy of 511-keV. The reality of high-energy emitters is that the
radiation exposure coming from the patient is much higher for PET than nuclear medicine.
Consequently, the room for error is smaller and some concerns that might be taken for
granted in nuclear medicine cannot be overlooked in PET or the tech will be overexposed.
Simple measures can minimize techs exposure time. Holbrook avers, Its
important to reiterate that the patient is where the tech can receive the most
exposure. Classic agrees. She adds, Lets face it. The patients are most
technologists primary source of exposure. Reducing time spent close to patients
reduces dose on the basis of time and distance.
This workstation is used for the preparation of PET
pharmaceuticals. Capintecs workstation can be used instead of a shielded hood or hot
cell.
In all hospitals, a radioisotope is carefully shielded before it is injected into a
patient in order to prevent accidental exposure. Once the isotope is injected into the
patient, it becomes an unshielded source of radiation. So a tech cannot opt to sit next to
a patient for reassurance after injecting radioisotopes, or he or she will receive an
undesirable exposure.
A typical PET scan requires anywhere from 30 minutes to 90 minutes of uptake time after
injection. A tech can be potentially exposed during this span of time. The patient should
be isolated in a shielded area during the uptake phase.
Nevertheless, a few sites designed their PET facility incorrectly and failed to account
for technologists safety during the uptake phase. Jeff Clanton, DPH, director of
radiopharmacy services and manager of the cyclotron facility for Vanderbilt University
(Nashville, Tenn.), notes, Ive seen a few PET facilities that arent
designed correctly. A PET site should have an isolated holding area for patients during
the uptake phase. The holding area should be away from staff and other patients.
The
Cyclotron Question
As hospitals look seriously at PET scanners or consider their
current PET situation they may also toy with the idea of investing in a cyclotron to
produce radiopharmaceuticals. Jeff Clanton, DPH, director of radiopharmacy services and
manager of the cyclotron facility for Vanderbilt University (Nashville, Tenn.), says,
The primary reason facilities want to purchase a cyclotron is to produce the newer
tracers that have a shorter half-life than FDG. These tracers include 15O, 13N, and
11C, which have half-lives of two minutes, 10 minutes and 20 minutes, respectively.
Most of the new small PET cyclotrons are self-shielded, so
hospitals dont need to build a lead vault to house the cyclotron. In fact, Clanton
notes, The shielding is excellent on most new machines with readings anywhere in the
room below 2 mrem. Cyclotron safety, however, does not end with shielding. It is
important to take into account activities in adjacent rooms. Siting a shielded cyclotron
in a room next to a patient waiting room, billing clerk work area, or any room with a high
occupancy of non-radiation personnel should be avoided.
Other radiation safety considerations with self-shielded
cyclotrons are the preparation area and the handling of large activities of positron
emitters. Remote manipulators, heavily shielded glove boxes and robotics that can do much
of the prep work reduce staff radiation dose considerably.
An unshielded cyclotron requires a 6-foot thick concrete
vault with a lengthy maze for neutron exposure reduction. The cyclotron manufacturer can
help design the vault. Despite the importance of radiation safety, it may be less of a
concern than other issues namely staffing. Clanton concludes, The biggest
factor with a cyclotron is hiring trained personnel to operate the equipment. It is a huge
cost; the hospital either needs to steal someone from another facility or train someone
on-site. |
While a hospitals design may overlook safety during the uptake phase, individual
technologists may slip into poor habits. Holbrook explains, If a technologist
injects a patient first and then starts asking questions and filling out forms, he will
receive some exposure from the patient. Another procedural no-no is to bring the
radiopharmaceutical into the room prior to starting the IV. The tech should start the
patients IV before bringing the radiopharmaceutical into the room. Glenn Sullivan,
radiation safety officer for Childrens Memorial Hospital (Chicago), outlines the
basic time and distance safety procedures for PET scanning.
Bring the patient into the room
Take the patients vital signs
Prepare the radiopharmaceutical dose
Calculate the specific activity needed
Complete the scan, standing 10-12 feet away at the computer during the scan
Classic adds a simple guideline for PET techs to keep in mind: techs should stand a
minimum of one to three feet away from the patient whenever possible.
Another potential PET safety concern is extremity exposures for staff members who prep
and administer the radiopharmaceuticals. Shielded PET syringes do help reduce the dose to
the body and hands. Clanton notes, With PET radiopharmaceuticals you need to have
the right PET syringe shields. The shields for standard nuclear medicine
radiopharmaceuticals just dont work.
PETNet Pharmaceuticals tracks maintenance and service of one of
its cyclotrons where FDG is produced.
Syringe shielding is just one part of the PET shielding battle. PET scanners require
more lead shielding than standard diagnostic radiology rooms. For most diagnostic
radiology rooms, 1/16 to 1/8 inch lead shielding suffices.
PET shielding requirements are calculated based in the size of the room, throughput of
patients and types of procedures. Classic says a few manufacturers continue to insist that
standard nuclear medicine shielding will suffice for a PET room; however, when most
institutions complete these calculations they find that additional shielding. Mayo Clinic,
for example, has 3/4 inch lead on its PET scan room walls. Some PET facilities also opt to
invest in portable lead shielding for the technologist to stand behind if he or she has to
be the in the room with the patient for a long time. Another option is to install lead
shielding on the wall between the scanner controls and the scan room.
Some sites may be considering adding a PET-CT scanner in addition to or instead of a
standard PET scanner. And while there are a variety of factors to weigh in the decision,
shielding is not one of them. Classic explains, The good thing about PET-CT is if
you plan for PET shielding, the CT part is taken care of. The shielding requirements for
PET easily cover CT.
Monitoring is a key piece of PET radiation safety. Holbrook says, Its a
good idea for techs to get radiation dosimetry badge readings on a regular monthly basis.
If someone is doing something wrong you dont want to wait two months to find about
it.
Although PET isotopes have a higher energy than nuclear medicine isotopes, they also
have a shorter half-life. Standard nuclear medicine isotopes have half-lives ranging from
six hours to one to two days, while PET isotopes have half-lives in the 15 seconds to 100
minute range. As a result, spill protocols differ. Typically, nuclear medicine spills are
handled immediately. If a PET isotope is spilled, allowing the material to decay is an
option because it decays relatively quickly. Holbrook says, In the event of a spill,
you may be better off covering the spill with lead shielding, closing off the area and
cleaning the spill later.
PET does have unique safety considerations; however, they are not at all
insurmountable. Holbrook concludes, The key to PET radiation safety success is
finding good protocols and being very consistent with them. With training, a good nuclear
medicine technologist can adapt to PET very easily.
Regulatory requirements
It seems as if everyone has his or her fingers into the PET regulations. The
Joint Commission on Accreditation of Healthcare Organizations (JCAHO of Oakbrook Terrace,
Ill.), the U.S. Nuclear Regulatory Commission (NRC of Washington, D.C) and individual
states all have some say in accreditation. And the really gung-ho may can explore
voluntary PET accreditation through the Intersocietal Commission for the Accreditation of
Nuclear Medicine Laboratories (ICANL of Columbia, Md.), which offers accreditation in PET
imaging as well as nuclear medicine and nuclear cardiology.
JCAHOs primary concern is general safety measures, so a JCAHO site visit may not
be intensely focused on radiation safety. When it comes to PET and nuclear medicine, JCAHO
generally looks at many of the same things it does in other areas of the hospital: proper
management of the patient, medications handling, infection control issues, and fire and
life safety. During a site visit, JCAHO may review employee exposure records and make sure
that the radiation safety committee is meeting regularly. They also may look at staff
training, orientation and certification records, and equipment calibration records.
NRC, on the other hand, is wholly focused on regulatory compliance and what a hospital
does to comply with regulations. NRC does take an in-depth look at radiation safety
issues, including radiation records, training and events, exposure records and radiation
safety committee records. Classic notes, NRC tends to be much more detailed than
JCAHO [when it comes to radiation safety]. NRC seems to be changing. It used to be very
paper-intensive, but during our last visit, they were more concerned with observing staff
and watching procedures. It was more interactive than in the past.
Finally, ICANL touts a number of benefits to its voluntary PET and nuclear medicine
accreditation program. According to the organization, accreditation demonstrates a
commitment to patient care and accountability, provides a confidential peer-review, and
can be used as a recruiting tool. Regardless of the decision to pursue accreditation,
however, radiation safety warrants a front seat when it comes to PET implementation.