More on DQE
I commend Medical Imaging on taking on the difficult but important subject of
quantitatively measuring image quality with digital detectors (MTF, DQE and the
Great Digital Detector Debate, May 2001, p. 72). The insight, along with the
comments of Dr. Dobbins from Duke University Medical Center and Dr. Maidment of Thomas
Jefferson University Hospital, was very refreshing.
I want to expand on two of the points raised in the article, namely that DQE is not a
point, but a curve and the speed/dose trade-off.
DQE is not a point but a curve. Most of the manufacturing literature and some of the
scientific literature leads the reader to believe that measurements of image quality like
Modulation Transfer Function (MTF) and Detective Quantum Efficiency (DQE) produce a single
score. Even the article implies the same. In the figure titled
Comparison of Detector Image Quality the text states that Direct
detectors generally have better MTF, but the a-Se direct detectors generally stop fewer
X-rays than the indirect CsI:Tl detectors, thus they have lower DQE. This begs the
question of lower DQE, or better MTF, at what spatial frequency?
When comparing the performance of a system using MTF or DQE measures, it is important
to consider the relevant portions of the DQE and MTF curves and not just a single spatial
frequency such as 0 cycles/mm.
As Dr. Dobbins and Dr. Maidment point out, the finer the structure, the higher the
resolution needed for faithful image representations. The diagnostic range for small bones
is between 1 and 2.5 cycles/mm, while the diagnostic range for examining the intestine is
between 0.3 and 0.6 cycles/mm. In general, the diagnostic range for general radiographic
imaging is between 0.5 and 3.0 cycles/mm. This is the portion of the MTF or DQE curve
where detector merit is most important and this is where the comparison between different
systems should be made
While it is true that at zero spatial frequency the current versions of a-Se detectors
have lower DQE than some CsI:Tl systems, the curves for these systems cross and the
selenium detector exhibits both higher MTF and higher DQE for spatial frequencies in the
areas of greatest interest to most radiologists.
Continuing in this same vein, there is no technical limitation prohibiting selenium
detector manufacturers from depositing selenium in a sufficiently thick layer so as to
have higher x-ray absorption than CsI:Tl and exhibit higher DQE even at zero spatial
frequency. By comparison, making CsI:Tl thicker for indirect detection detectors is not
useful, because the DQE would fall off rapidly at the higher spatial frequencies.
Speed/Dose Trade-Off. Detectors designed for digital mammography highlight the
speed/dose performance tradeoff issue inherent in systems that use scintillators like
CsI-based systems. These systems must have a sufficiently thick CsI layer so that the
speed of the systems is not inferior to screen-film mammography. However, scintillator
systems exhibit relatively poor resolution compared to screen-film. In comparison,
selenium can be easily made essentially 100 percent absorbing at the low mammographic
energies, and result in a digital detector having superior DQE (and MTF) over their entire
spatial frequency range (from 0 to Nyquist) compared to either screen-film or CsI:Tl
systems.
Andrew Smith, Ph.D.
Imaging Sciences
Hologic, Inc.
| Send your letters (200 words maximum) to: Editor,
Medical Imaging, 295 Promenade St., Suite 2, Providence, RI 02908-5720. Fax:
401-455-1555. Or, via e-mail at mtierney@mwc.com. All letters are subject to editing for
both space and clarity. |