AAMI's task force support welcomed
I am writing to express concern about an article entitled "AAMI-sponsored
regulation group submits to FDA." The article appeared on page 29 in the December
1999 issue. I have received several phone inquiries that indicate that your readers have
been misled by the article.
The agency's strategy for minimal regulatory controls
over the remarketing and servicing industries is based on cooperation between the various
stakeholders and FDA. The AAMI proposal established a commitment by those organizations
that participated in the Industry Task Force to develop an effective voluntary program for
self-regulation. Accordingly, we plan to issue a guidance document that will probably
incorporate a significant portion of the AAMI proposal. After publication of the guidance
document, we will evaluate the need for further federal regulation beyond what is already
required.
The agency appreciates the willingness of the AAMI
Industry Task Force to develop a voluntary approach to the complex issue of remarketed
medical devices. More importantly, we appreciate the merits of the AAMI Industry Task
Force proposal, and we will use it accordingly.
Walter W. Morgenstern
Deputy Director, Division of Program Operations (HFZ-305)
Office of Compliance
Setting the record straight on QS regs
The article "Pre-Owned Equipment ... When it's the Right Choice," in
the February 2000 issue touches upon a number of interesting and pertinent matters.
However, unless this article was written five years ago, the information under the
subheading FDA guidelines is incorrect. The new GMP, now called the Quality System (QS)
Regulation, was published in final form Oct. 7, 1996. There is no revision pending, nor
draft that was to have been issued in the month of February 2000. Medical device GMPs have
regulated the industry since 1978, not 1987 as the article states.
More importantly, refurbishers can be divided into two
categories: medical device manufacturers who refurbish their own devices, and third-party
refurbishers who recondition original equipment manufacturers (OEMs) medical devices. This
article seems to be talking more to the third-party refurbishers, but as far as the
regulatory requirements are concerned, it does not matter. All of the types of medical
equipment to which this article refers are medical devices according to section 201(h) of
the Food, Drug & Cosmetic Act. They are therefore subject to the laws and regulations
governing medical devices.
Accordingly, and in accordance with the QS regulation, all
refurbishers of medical devices are considered manufacturers, and are required to comply
with all provisions of the QS regulation, as well as many other applicable regulations.
Compliance is not voluntary, as this article states. If a third-party refurbisher takes it
upon himself/herself to modify the original equipment design, he/she would be obligated to
follow the design control requirements of the QS regulation.
If the engineering design is not modified, the third-party
refurbisher would be obligated to verify and validate that the refurbished equipment meets
all of its original safety, effectiveness and performance specifications.
Regarding any claims the third-party refurbisher makes
about the reconditioned equipment, he/she would be held responsible to provide
scientifically valid evidence supporting those claims.
This section of the article gives an incorrect and
distorted impression of the regulatory obligations imposed on refurbishers, and I think it
would be important to publish an appropriate correction.
Michael Zmuda, Ph.D., RAC
Principal, Z-Tecknologie, S.P.
Send your letters (200 words maximum) to: Editor,
Medical Imaging, 295 Promenade Street, Suite 2, Providence, RI 02908-5720. Fax
401-455-0555. Or, via e-mail at mtierney@medworldcomm.com.
All letters are subject to editing for both space and clarity. |